NEX CG II
Energy Dispersive X-ray Fluorescence Spectrometer
Next-generation Advanced Cartesian Geometry EDXRF for Rapid Qualitative and Quantitative Elemental AnalysisNEX CG II, a powerful second-generation energy dispersive X-ray fluorescence (EDXRF) spectrometer, delivers rapid qualitative and quantitative determination of major and minor atomic elements in a wide variety of sample types — from oils and liquids to solids, metals, polymers, powders, pastes, coatings, and thin films.
- Non-destructive elemental analysis for sodium (Na) to uranium (U)
- Quick elemental analyses of solids, liquids, powders, coatings, and thin films
- Indirect excitation for exceptionally low detection limits
- High-power 50 kV 50 W X-ray tube
- Large-area high-throughput silicon drift detector (SDD)
- Analysis in air, helium, or vacuum
- Powerful and easy to use QuantEZ® software with multilingual user interface
- Advanced RPF-SQX Fundamentals Parameters software featuring Scattering FP
- Rigaku Profile Fitting (RPF) advanced algorithm for peak deconvolution
- Various automatic sample changers accommodating up to 52 mm samples
- Low cost of ownership backed by a 2-year warranty
Industrial Quality Control to Advanced Research Applications
The Rigaku NEX CG II is a multi-element, multi-purpose analyzer, ideal for measuring ultra-low and trace element concentrations into the percent levels. Especially well-suited for semi-quantitative determination of elemental content in complete unknowns, NEX CG II serves many industries. Applications range from research and development to industrial and in-plant quality assurance. It is easy to use for non-technical operators yet powerful enough for expert use in commercial labs and R&D facilities. Users can achieve the lowest limits of detection and easily manage complex applications like testing agricultural soils, and plant materials, analyzing finished animal feeds, measuring waste oils, environmental monitoring, and many others.
Cartesian Geometry and Polarization for Trace Level Sensitivity
Unlike conventional EDXRF spectrometers, NEX CG II is an indirect excitation system using secondary targets rather than tube filters. Monochromatic and polarized excitation from secondary targets vastly improves detection limits for elements in highly scattering matrices like water, hydrocarbons, and biological materials. Secondary target excitation in full 90° Cartesian Geometry eliminates background noise. As a result, NEX CG II brings a new level of analytical sensitivity to XRF technology. Users can measure ultra-low and trace element concentrations, even in challenging sample types.
Easy Instrument Control with Advanced Qualitative and Quantitative Analytical Software
NEX CG II is easy to use with QuantEZ, a powerful PC-based software providing intuitive instrument control with simple menu navigation and a customizable EZ Analysis interface. Users can maximize their time and productivity with simplified routine operations and create their own methods using a simple flow bar wizard.
Advanced qualitative and quantitative analysis is powered by Rigaku’s RPF-SQX Fundamental Parameters (FP) software, featuring Rigaku Profile Fitting (RPF) technology and Scattering FP.
Technical Data
Excitation
- X-ray tube, end-window type with Pd anode
- 50 W max power
- 50 kV max voltage
- Five standard polarization and secondary targets for optimum excitation
Detection
- Large-area high-throughput silicon drift detector (SDD)
- Peltier electronic cooling
- Digital pulse processor
- Automated or user confi gurable shaping times for optimum analytical performance
Sample chamber
- Large 32.5 cm diameter x 7.5 cm deep sample chamber allows for various sample sizes
- Analysis in air
Environmental Conditions
- Ambient temperature 18 – 28°C (65 – 82°F)
- Relative humidity ≤75%
- Vibration undetectable by human
- Free from corrosive gas, dust and particles
Software
- QuantEZ® software for control of spectrometer functions and data analysis
- Simple flow bar wizard to create your own methods
- Empirical calibration with overlap and matrix compensation
Computer
- External PC computer system
Single phase AC 100/220 V, 3.8/1.6 A (50/60 Hz)
Dimensions: 46.3 (W) x 49.2 (D) x 38.2 (H) cm (18.2 x 19.4 x 15.0 in)
Weight: 65.8 kg (145 lbs.)
Applications
- Sulfur in oil - ASTM D4294, IP 336, ISO 8745, ISO 20847, JIS K 2541-4, ISO 13032, ASTM D7220, EPA ULSD, EPA Tier 3
- Lead in Gasoline - ASTM D5059
- Chlorine in Crude Oil - ASTM D4929
- Lube oils - ASTM D6481, ASTM 7751
- Wood Treatment
- Phosphorus Fire Retardant on Fabric
- Silicone Coating on Paper/Plastic
- RoHS
- Cement including ASTM C114
- Coatings
- Mining & refining
- Metals and alloys
- Paint and pigments
- Education
- Catalysts
- Wovens and non-wovens
Inquiry - NEX CG II Energy Dispersive X-ray Fluorescence Spectrometer